Environmental Conditions
|
|
For operation |
For storage |
|
Temperature |
+5degC to +35degC |
-5degC to +50degC |
|
Atmospheric pressure |
850 to 1070 hPa |
550 to 1100 hPa |
|
Humidity |
0 to 65% r.H |
0 to 65% r.H. |
Performance data
Flow of free air
The free air delivered by the pump at various stroke rates is as follows:-
|
Strokes/min |
Free air delivered (litres/min) |
|
30 |
6.3 litres/min |
|
40 |
7.7 litres/min |
|
50 |
9.4 litres/min |
Performance under load
The flow and average pressure against an Amal jet no.45 at various stroke rates at 25degC are as follows:-
|
Strokes/min |
Free air delivered |
Average Pressure |
Peak Pressure |
|
(litres/min) |
upstream of Jet (kPa) |
upstream of Jet (kPa) | |
|
20 |
3.7 |
36 |
69 |
|
30 |
5.2 |
68 |
115 |
|
40 |
6.2 |
98 |
144 |
|
50 |
7.1 |
127 |
175 |
|
50* |
6.0 |
102 |
139 |
* at 35degC and 65% relative humidity.
The stroke rate required to maintain a flow of 4 litres against resistance of Amal jet No. 45 is 23 strokes/min at 25degC. The work required per stroke is 80N at a stroke rate of 30/min. Maximum recommended stroke rate, 50 strokes/min.
Effect of ambient temperature
The volume of free air delivered by the pump and the average and peak pressures will decrease with increasing temperature. Typical results at 25degC & 35degC are given in the table above.
Particle filter
Filter structure: polyamide monofilament, woven fabric, mesh aperture 10 micro; disposable; maximum life 12 months.
Weight: 700g
Dimensions (h x w x d).
Packed 111 x 190 x 306 mm.